Stay Informed
Follow us on social media accounts to stay up to date with REHVA actualities
Guangyu CaoNorwegian University of
Science and Technology, Trondheim, Norway | Anders Mostrøm NilssenNorwegian University of
Science and Technology, Trondheim, Norway | Hans Martin MathisenNorwegian University of
Science and Technology, Trondheim, Norway | Yixian ZhangChongqing University, China |
Kai XueChongqing University, China | Liv-Inger
StenstadSt.
Olavs hospital, Norway | Andreas RadtkeSt. Olavs hospital, Norway | Jan Gunnar SkogåsSt. Olavs hospital, Norway |
At present,
laminar airflow (LAF) systems and mixing ventilation (MV) systems are two
commonly used ventilation solutions for operating rooms (ORs) to ensure the
required indoor air quality. Recent studies have shown that there is little
difference in the prevalence of surgical site infection (SSI) between LAF
systems and MV systems. The objective of this study was to compare the
performance of a LAF system and a MV system in ORs at St. Olavs hospital,
Norway. In this study, all experimental measurements were conducted in real ORs
at St. Olavs hospital in Trondheim, Norway. The results showed a wide range of air
distribution patterns in the surgical microenvironment with both systems. Under
operating conditions, the thermal plume from a lying patient and surgical staff
may change the local airflow distribution in the surgical microenvironment in
the OR with LAF. This indicates that MV may be a robust way to deliver airflow
under disturbed conditions. This study suggests that the performance of LAF and
MV needs to be evaluated regularly under real surgical procedures in Norwegian
hospitals.
Surgical
site infection (SSI), which are the most common hospital-acquired infections,
leads to a big burden for the patient and an increased cost for the society.
Among other factors, the air quality of operating rooms (ORs), especially the
surgical microenvironment (see Figure 1), plays an important role to
prevent the development of surgical site infections (SSIs). One previous study
shows that an improved indoor environment of a hospital building can reduce
costs associated with airborne illnesses by 9% – 20% [1]. At present,
both laminar airflow (LAF) systems and mixing ventilation (MV) systems are commonly
used in ORs to ensure the required indoor air quality. Figure 1 shows sketches of an operating theatre with a mixing system and a
laminar airflow system.
Figure 1.
Principle of ventilation systems in operating rooms: a) a vertical laminar
system, b) a mixing ventilation system. [2]
Recent
studies have shown that there is little difference in the prevalence of SSI
between the designs of an LAF system and an MV system. The recently published
WHO guideline suggests that LAF systems should not be used to reduce the risk
of SSI for patients undergoing total arthroplasty surgery, but the conclusion
is disputed and based on conditional recommendation, low to very low quality of
evidence [3]. In fact, ORs contains numerous transient phenomena that may cause
significant changes to the time resolved indoor air distribution patterns.
Multiple studies have investigated how different factors affect the efficiency
of the two different ventilation systems. Table 1 summarizes these findings.
Table 1. A comparison of LAF and MV. [4]
Aspects | LAF
| MV
|
Position of the operation table and the sterile
operating team | Very important. Has specific borders between the
sterile zone and the surroundings. | Not so important. Designed to provide equal
conditions in the entire room. |
Type and position of the lamps | Very important [5]. It was
identified that the positioning of lamps is crucial to the airflow
distribution near the patient. | Less important. Mixing airflow will dilute the
contamination concentration in the whole operating room. |
Operating staff clothing system | Very important. To great extent determines staff
source strength. | Very important. To great extent determines staff
source strength [6]. |
The reason
for these controversial results and conflicting guidance is the lack of
scientific understanding of the dynamic distribution in the surgical
microenvironment (see Figure 1) in ORs under operating conditions.
The objective of this study was to compare the performance of LAF systems and
MV systems in terms of airflow distribution in the surgical microenvironment in
ORs at St. Olavs hospital.
In this
study, all measurements were conducted in two ORs at St. Olavs hospital in
Trondheim, Norway. The OR with an LAF has an area of 56 m² with 11 m²
of laminar airflow zone, which is surrounded by 1.1 m long partial walls
(see Figure 2). During the experimental
measurements, the ventilation system was operated with full load, and the room
temperature was commonly set to 22°C. During the experiments, the supply
air temperature was 20 ±1°C. The designed supply air in the
orthopedic OR with LAF was 10 580 m³/h: comprising 4 280 m³/h
of outdoor air and 6 300 m³/h of recirculated air. A male thermal manikin was used to simulate a patient
in an operating room. The detailed description of the thermal manikin can be
found in Cao et al. (2018) [7].
The OR with
an MV system was equipped with four ceiling-mounted diffusers. For the exhaust,
there were two wall-mounted exhaust outlets and one near the ceiling. The OR
with MV had an area of 59.7 m². The set-point temperature of the theatre
was 22.0°C in all scenarios. The supply airflow rate was 3 700 m³/h, and
the exhaust airflow was 3 600 m³/h. During measurement, an adjustable
stand was used to carry the anemometers. Five anemometers were aligned on the
stand with a separation of 10 cm. The stand was placed at three different
positions above the operating table: pelvis, waist and chest. At each
cross-section, measurements were performed at six heights: 5, 10, 15, 20, 25,
and 30 cm above the surface of the location. The heights of the
measurement point were selected to present relative to the human body, which
does not have equal heights at each part of the body surface.
a) | b) | c) |
Figure 2.
Experimental setup: a) photo of the operating room with a LAF; b) photo
of the operating room with an MV; c) location of measurements. [4]
In this
study, two scenarios (see Table 2) that include four different cases,
were investigated. Scenario 1 (cases 1-2) measured the airflow
distribution in these ORs with only an operating table as a reference case.
Scenario 2 (cases 3-4) measured the airflow distribution in the ORs with a
lying patient. Operating lamps were put away from the measurement zone.
Table 2. Scenarios of the experimental measurements.
Scenarios | Cases | Number
of patients | Ventilation
mode |
S1 | case 1 | 0 | LAF |
case 2 | 0 | MV | |
S2 | case 3 | 1 | LAF |
case 4 | 1 | MV |
The
AirDistSys 5000 system with five omnidirectional anemometers was used to
measure the velocity and temperature of the airflow near the operating table.
The velocity range of the SensoAnemo 5100 LSF omnidirectional anemometers is
0.05 – 5.00 m/s with an accuracy of ±0.02 m/s ±1.5% of
readings. The recording time for each measurement row was set to 3 minutes.
Figures 3 a-d show the velocity distribution
above an empty operating table in ORs with LAF and MV. Figures 3a and 3b show the velocity contours above
the chest of the patient in ORs with LAF and MV, respectively. With the LAF
system, the velocity above the chest position is 0.15 – 0.26 m/s,
which is similar to the velocity distribution with the MV. The velocity
contours in the LAF system show a downward airflow pattern, and the velocity
contours in the OR with MV shows a side-blow (from left to right) airflow
pattern. Figures 3c and 3d
show the velocity above the waist position in two ORs with the LAF and the MV,
respectively. In the OR with LAF, the minimum value is 0.18 m/s, and the
maximum is 0.32 m/s. The results show that the velocity distribution
varies in these two systems. The airflow distribution in the OR with LAF
resembles a stratified airflow with decreasing velocity when it approaches the
operating table. The velocity distribution in the MV system is more similar at
different positions.
a) | b) |
c) | d) |
Figure 3.
Measured velocity contours above operating table in ORs with LAF and MV –
scenario S1 including case 1 and case 2: a) above the chest position with
an LAF system; b) above the chest position with an MV system; c) above
the waist position with an LAF system; d) above the waist position with an
MV system.
Figures 4 a-d show the velocity distribution above a lying patient in ORs with LAF and MV. Figure 4a and 4bshow the velocity contours above the chest of the patient in ORs with LAF and MV, respectively. With the LAF system in Figure 4a, the velocity above the chest position was 0.12 – 0.24 m/s. The velocity near the patient was notably low (0.12 m/s) because of the thermal plume generated by the patient. Figure 4b shows a similar distribution with the MV system, which generated a slightly higher velocity zone (0.16 m/s) notably near the chest. Figure 4c shows the velocity distribution above the waist in the OR with LAF. It shows that the velocity near the patient became even lower above the waist, 0.08 m/s. The plume-like airflow distribution may be caused by the rising thermal plume from the patient. As Figure 4d shows, the velocity measured above the waist varies between 0.14 – 0.20 m/s, which is similar to that in Figure 4b, which was measured above the chest.
a) | b) |
c) | d) |
Figure 4.
Velocity contours above a lying patient in ORs with LAF and MV – scenario S2
including cases 3 and 4: a) above the chest position with an LAF system; b) above
the chest position with an MV system; c) above the waist position with an
LAF system; d) above the waist position with an MV system.
This study only presents the measurement results in ORs without the use of surgical lamps, which may affect the airflow pattern significantly. One earlier study reported the measured air velocity profiles formed under the surgical lights and without lights for different heights [7] (shown in Figure 5). The edges of lights are highlighted with dashed lines of the same matching color that is used for the velocity profiles. The turbulent airflow formed behind lights is illustrated by the gap formed between the yellow marked points representing velocities measured without lights and the points measured under different surgical lights – marked by blue, red and green colors. The mean value of the velocities measured under surgical lights were 0.07 m/s under light mo. 1 (blue), 0.07 m/s for light mo. 2 (red) and 0.06 m/s for light mo. 3 (green). The measured air velocity was over 0.25 m/s without the effect of surgical lamps.
Figure 5. Measured mean velocity profiles
with and without the effect of surgical lamps at different heights in ORs. [8]
In addition
to airflow distribution, another previous study investigated the turbulence
intensity in the surgical microenvironment with an MV and an LAF [4]. Figure 6 a-b show the measured air turbulence intensity distributions above a lying
patient surrounded by three surgical staff with the use of two surgical lamps. Figure 6a shows measured turbulence intensity in an OR with an LAF. The values
range from 5 to 20% at 15 cm above the body surface. While the highest
values, 20 to 25% are encountered within 10 cm from the body surface. Figure 6b shows the measured contours of air turbulence intensity in the
operating room with MV, which varies from 30% to 40% above the waist of the
simulated patient. These results indicate that air turbulence intensity level
of supply airflow from LAF is much lower than from MV. This was caused by the
mixing processed of supply air and ambient air in operating rooms in the
surgical microenvironment.
a) | b) |
Figure 6. Air turbulence intensity contours above a lying patient surrounded by three surgical staff with the use of two surgical lamps: a) above the waist position with an LAF system; b) above the waist position with an MV system. [4]
The air
distribution in operating rooms may significantly change under real operation
conditions with various disturbance, including surgical facilities, internal
heat sources, patients, surgical staff and various monitors. A common feature of
the airflow pattern in ORs with either LAF or MV is that the velocity contours
are drastically changed from each cross-section, which indicates the combined
effect of surgical facilities and the thermal plume of the patient and surgical
staff. However, the surgical lamps appear to have a greater effect on the velocity
with an LAF system than with an MV system. This study provides evidence that
the airflow velocity in the surgical microenvironment shows a wide range of patterns
with an LAF system and an MV. The thermal plume from a lying patient may change
the airflow distribution in the surgical microenvironment more in the OR with an
LAF than with an MV system. This study suggests that the performance of LAF and
MV need to be evaluated under different surgical procedures in Norwegian
hospitals. Further studies are needed to clarify how these different airflow
patterns will influence the development of SSIs. More experiments using tracer
gas need to be performed to investigate the effect of MV and LAF on the heat
and mass transfer in the surgical microenvironment.
The authors
greatly appreciate the collaboration with the Operating Room of The Future
(FOR) – St. Olavs hospital, which provided real operating rooms for field
measurements. Norwegian University of Science and Technology provided in-kind
funding to support the field measurements at St. Olavs hospital.
[1] A. Shajahan, C.H. Culp, B. Williamson.
Effects of indoor environmental parameters related to building heating,
ventilation, and air conditioning systems on patients' medical outcomes: A
review of scientific research on hospital buildings. Indoor Air 29 (2019)
161-176.
[2] A.M. Nilssen, Characterization of the
airflow distribution in close proximity to the patient in an operating room.
Master thesis, Norwegian University of Science and Technology. 2018.
[3] WHO Guideline, Global guidelines for the
prevention of surgical site infection, World Health Organization 2016.
[4] G.Y. Cao, A. M. Nilssen, Z. Cheng, L.I.
Stenstad, A. Radtke, J.G. Skogås. Laminar airflow ventilation and mixing ventilation: which is better for
operating rooms regarding airflow distribution near an orthopaedic surgical
patient? American Journal of Infection Control 47 (2019) 737–743.
[5] H. Brohus, K. Balling, D. Jeppesen. Influence of movements on
contaminant transport in an operating room. Indoor Air 16 (2006) 356-372.
[6] A. Tammelin, B. Ljungqvist, B. Reinm€uller. Comparison of three distinct
surgical clothing systems for protection from air-borne bacteria: a prospective
observational study. Patient Safety in Surgery 6 (2012) 23.
[7] G.Y. Cao, M. Storås, A. Aganovic, L.I.
Stenstad, J.G. Skogås. Do
surgeons and surgical facilities disturb the clean air distribution close to a
surgical patient in an orthopedic operating room with laminar airflow? American
Journal of Infection Control 46 (2018) 1115-1122.
[8] A. Aganovic, G.Y. Cao, L.I. Stenstad. J.G. Skogås. Impact of surgical lights on the velocity distribution and airborne contamination level in an operating room with laminar airflow. Building and Environment 126 (2017) 42-53.
Follow us on social media accounts to stay up to date with REHVA actualities
0